Official recall record
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedi
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-07-30
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 07613327033847
- Brand on record
- Howmedica Osteonics Corp.
What the authority reported
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-15.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2023-12-01 · US FDA — Medical Devices
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
Other recalls from this brand
- 2026-09-16 · US FDA — Medical Devices
TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507. - 2026-09-16 · US FDA — Medical Devices
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519. - 2026-09-16 · US FDA — Medical Devices
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.
All 20 official recall notices naming Howmedica Osteonics Corp. on this site →