DIVA 24 Inch Widescreen LCD Touch Display - Used with the Philips PIC iX system central monitoring system. Model: RGD2461AMI Philips Pa

US FDA — Medical Devices · 2024-02-15 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-02-15
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Philips North America Llc

What the authority reported

If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a patient condition which prompted the alarm to generate. Although unlikely, this could potentially result in a patient harm.