Official recall record
Intrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2026-08-18
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00849436000020
- Brand on record
- Percussionaire Corporation
What the authority reported
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-22.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2026-08-18 · US FDA — Medical Devices
Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Inst
Other recalls from this brand
- 2026-08-18 · US FDA — Medical Devices
Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual - 2026-08-18 · US FDA — Medical Devices
Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC - 2026-08-18 · US FDA — Medical Devices
Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used