ProductGuru Recalls

Official recall record

AK 98, 115V BIO/ AK 98, 115V SELF-CARE and AK98 V2 115V BIO VERSION / AK 98 V2 115V BIO SELF CARE/ AK 98 V3 (2018-08-29)

Health Canada — Medical Devices · 2021-10-31 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2021-10-31
Last updated (per notice)
2024-03-26
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Baxter is communicating important safety information regarding the potential for AK 98 hemodialysis devices to generate excessive Ultrafiltration (UF) in certain situations where treatment-related alarms occur, or where there is an ultrafilter leak. Excessive UF may present hazards for sensitive patients, such as low-weight patients, for whom target UF values of zero or very low volumes are desired. Recall start date: August 29, 2018

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.