Official recall record
Astral Ventilators
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-10-13
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Resmed's Astral ventilator is designed to deliver a Total Power Failure (TPF) audible alarm for at least 2 minutes to alert the carer/clinician when the ventilator experiences a total loss of power that results in a stop of ventilation. The TPF audible alarm is powered by an electrical component known as a supercapacitor. This component degrades over time, which may cause the TPF alarm to sound for less than 2 minutes, or in some cases, to not sound at all. Recall start date: September 13, 2023