ProductGuru Recalls

Official recall record

Skytron Freedom Boom

Health Canada — Medical Devices · 2023-10-18 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2023-10-18
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

The manufacturer received three reports that during processing and preparation of the room setup the monitor suspension disconnected from the height adjustable arm construction. This affects all lots of the lightweight spring arm (part numbers 42A-43L, 42A-43M, and 42A-43H) which are built into Freedom Boom models F110, F120, F320. Recall Start Date: October 10, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters