Official recall record
Skytron Freedom Boom
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-10-18
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
The manufacturer received three reports that during processing and preparation of the room setup the monitor suspension disconnected from the height adjustable arm construction. This affects all lots of the lightweight spring arm (part numbers 42A-43L, 42A-43M, and 42A-43H) which are built into Freedom Boom models F110, F120, F320. Recall Start Date: October 10, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.