Compress Anchor Plug

Health Canada — Medical Devices · 2023-10-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-10-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Zimmer Biomet is conducting a lot specific medical device recall for the compress device segmental anchor plug and the compress device short anchor plug products. The affected anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through. The issue was determined to be caused by a manufacturing process update impacting select lots manufactured prior to 2019. There have been five complaints received for the issue. The potential interference issue would be recognizable at the point of use. Any products previously implanted from the affected lot numbers are not impacted by this recall. Recall Start Date: October 16, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters