Official recall record
EPIQ Diagnostic Ultrasound System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-10-19
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips EPIQ Ultrasound systems operating with software version 10.0 may present an image lag, where the user is expecting to see a real time image and the system delays the presentation of that image by several seconds. If present, this issue will be observed in the following scenario: • The user begins to live scan and they do not see an image for several seconds (image is not immediately displayed). • The image lag will continue to persist (image displayed will continue to be several seconds behind). Recall Start Date: October 13, 2023