ProductGuru Recalls

Official recall record

EPIQ Diagnostic Ultrasound System

Health Canada — Medical Devices · 2023-10-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-10-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips EPIQ Ultrasound systems operating with software version 10.0 may present an image lag, where the user is expecting to see a real time image and the system delays the presentation of that image by several seconds. If present, this issue will be observed in the following scenario: • The user begins to live scan and they do not see an image for several seconds (image is not immediately displayed). • The image lag will continue to persist (image displayed will continue to be several seconds behind). Recall Start Date: October 13, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.