Official recall record
Alaris™ Syringe Pump MK4
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-10-30
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
BD is issuing a notification to inform customers that Cardinal Health branded Monoject™ syringes have not been validated for use with the BD Alaris™ infusion devices. BD has validated Covidien branded Monoject™ syringes for use with the BD Alaris™ infusion devices. Cardinal Health, rebranded its syringes from Covidien Monoject™ to Cardinal Health Monoject™ and retained the original Covidien SKU numbers. BD's investigation determined that there may differences in syringe dimensions associated with these changes that may impact BD Alaris infusion devices. Recall start date: October 12, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.