Philips Achieva and Intera Systems
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-11-06
- Last updated (per notice)
- 2025-06-04
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Philips has identified an issue where a specific component failure in the gradient coil of the affected 1.5T MR systems listed in this letter may act as a heat source with a potential to produce smoke and/or fire. If the component failure occurs, the user may observe the following: Repeated scan abort error messages due to gradient amplifier fault detection, described in section 4. Unusual noise from the system in the examination or technical room Smoke and/or fire from within the system Recall start date: October 25, 2023 Update (2025-06-04): Philips has identified additional MR systems (Intera 1.5T).