Artis Icono Ceiling

Health Canada — Medical Devices · 2023-11-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-06
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Artis Icono Ceiling systems are equipped with an automatic collision detection at locations where collisions with users or equipment might occur. If the issue occurs, the collision sensor around the x-ray tube housing shows the following unintended behavior: when the Artis Icono Ceiling system was switched off for a longer time (e.g. overnight) and switched on, this collision sensor is falsely signaling non-existing collisions for up to half an hour. Recall start date: October 19, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.