Official recall record
Alphenix INFX-8000v - Main Unit
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-11-08
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
It has been found that due to problems with the internal board of x-ray flat panel detector configured in the x-ray interventional systems, abnormal images such as horizontal stripes may be displayed and recovery probably could not be possible by restarting the system. Recall start date: October 23, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.