ProductGuru Recalls

Official recall record

Biofire® Filmarray® Torch System Base

Health Canada — Medical Devices · 2023-11-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

When turning on biofire torch Systems, arcing inside of the power switch may result in carbon build‐up on the switch contacts. The carbon build‐up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use. Recall Start Date: October 27, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.