Official recall record
Biofire® Filmarray® Torch System Base
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-11-09
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
When turning on biofire torch Systems, arcing inside of the power switch may result in carbon build‐up on the switch contacts. The carbon build‐up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use. Recall Start Date: October 27, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.