ProductGuru Recalls

Official recall record

Titan® Inflatable Penile Prosthesis

Health Canada — Medical Devices · 2023-11-10 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-10
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Pump bulbs are manufactured using 2 molds. A component on one of the two molds shifted during production. This resulted in a gradual build of material on the mold, which gradually decreased wall thickness of the pump. This occurred over the course of several weeks, culminating in a period of time (17 September 2022 to 02 December 2022) where pump bulbs were manufactured below the wall thickness specification. Recall Start Date: October 30, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.