epoc® NXS Host – 5553CA
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-11-17
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under specific conditions when the (automatic) delete blood tests feature is enabled. When the results are run on the epoc® NXS host and delete blood tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. Recall start date: October 24, 2023