HeartStart Intrepid Monitor/Defibrillator (ref 74608)

Health Canada — Medical Devices · 2023-11-17 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-17
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips has identified a potential safety and compliance issue: the HeartStart Intrepid Monitor/Defibrillator's service manual does not detail electrical safety test verification methods as required by International Electrotechnical Commission(IEC) requirements for class I devices. The current HeartStart Intrepid service manual does not detail IEC class I electrical safety test verification methods. The lack of this testing does not introduce a failure, but it may prevent a failure from being detected. This could occur when an IEC class I device is tested as a class II device after a servicing event. The issue was identified as the result of a service provider's inquiry. There have been no reports of patient harm. Recall start date: October 19, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.