ProductGuru Recalls

Official recall record

Stat Profile Prime® Plus Analyzer

Health Canada — Medical Devices · 2023-11-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Nova Biomedical became aware of a potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new bun/creatinine sensor cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional quality control practices. Recall start date: September 13, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.