Video Telescope - Endoeye HD II

Health Canada — Medical Devices · 2023-11-27 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-27
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Olympus has identified a total of 1003 complaints, including 372 adverse events (since September 2020 to August 2023) related to pink or green coloration of the image, including cases with reported delays of treatments and/or prolonged surgery. Olympus investigated the complaints and identified that the CCD chip that produces the image became damaged. The CCD chip is sensitive to heat and mechanical shocks. Olympus's risk management is taking this into account, and the instructions for use mention necessary steps to prepare the video telescope prior to usage. This medical device recall has been launched to remind customers to follow the steps in the instructions for use, especially to inspect the image prior to a clinical procedure, and always have a spare laparoscope available. Recall start date: November 8, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.