ProductGuru Recalls

Official recall record

Vantage Galan 3T (ref 74659)

Health Canada — Medical Devices · 2023-12-01 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-12-01
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

It has been determined that the terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed. Because of this, the cable became disconnected, and arcing occurred. Melting of the cable due to this arcing caused the cable securing parts below it to ignite, resulting in fire at the terminal block. Recall Start Date: November 14, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.