ProductGuru Recalls

Official recall record

Bard Mission Disposable Core Biopsy Instrument Kit

Health Canada — Medical Devices · 2023-11-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-11-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD continues to receive complaints for the product regarding mismatch between the coaxial and the needle in mission kit devices. Based on the event reported, the internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue. Recall start date: November 8, 2023