Cardiohelp-i

Health Canada — Medical Devices · 2023-12-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-12-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The quality issue under investigation is applicable to the "connection cable venous probe, l0.23m". The part causing the nonconformance under investigation is responsible for data transmission between the Cardiohelp-i and the venous probe. A noticeable high number of breaks in the insulation sheath and a shielding breach of the venous probe connection cable was observed. Recall Start Date: November 21, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters