Ambu aView 2 Advance
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-12-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Ambu has received information on two incidents where Ambu® aView™ 2 Advance caught fire when mounted on the vesa holder of the aCart™ compact, due to incorrect length screws used penetrating the lithium-ion battery of the device. Investigations have identified that the fire hazard was a direct consequence of a battery short-circuit, which, in turn, was triggered by the use of excessively long screws to mount the device. There is no instruction for mounting in the IFU accompanying the device. Recall Start Date: November 18, 2023