Synchromed II - Programmable Pump
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-12-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Medtronic recently identified that if the Synchromed II Pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field. In this case, drug delivery will only resume after performing a post MRI pump interrogation with the clinician programmer which will end telemetry mode. Recall Start Date: November 21, 2023