Health Canada — Medical Devices · 2023-12-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-12-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Medtronic recently identified that if the Synchromed II Pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field. In this case, drug delivery will only resume after performing a post MRI pump interrogation with the clinician programmer which will end telemetry mode. Recall Start Date: November 21, 2023