ProductGuru Recalls

Official recall record

LITe Decompression Tube Snake Arm

Health Canada — Medical Devices · 2023-12-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2023-12-06
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Stryker received one (1) complaint for a LITe Decompression Tube Snake Arm that did not fit on to the arm post shaft that mounts to the hospital bed. Subsequent investigation determined a portion of one lot of the LITe Decompression Tube Snake Arm was manufactured with the incorrect clamp component. As a result, the affected snake arms cannot attach to the arm post shaft. No adverse events have been reported for this issue. Recall start date: November 20, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters