Official recall record
Maquet Surgical Lights
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-12-07
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Maquet S.A.S. /Getinge has received a customer complaint regarding a fall of a light system in an operating room in Hong Kong (no injuries reported). Tnvestigation into this event demonstrated that the recommendation of maintenance and service were not performed as instructed in the current user and maintenance manuals. This notification communicates retrospectively current maintenance instructions and programs to customers. Recall start date: November 29, 2023