Brightview System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2023-12-22
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has become aware of a potential safety issue affecting Brightview systems that may present an extremity entrapment hazard to patients during a scan. While using pre-programmed motion (PPM) during an extrinsic quality assurance scan, a gap is created between the patient support and the detector. This gap presents a potential extremity entrapment hazard for patients while the system detectors and patient support are in motion. Recall Start Date: December 8, 2023