ProductGuru Recalls

Official recall record

Captus® 4000e Thyroid Uptake System, all models

Health Canada — Medical Devices · 2024-01-02 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-01-02
Last updated (per notice)
2024-04-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

In 2023, three Captus® 4000e systems incurred a spring arm failure, causing the arm to fall. The collimator and arm, together, weigh 45 lbs (20.4 kg), which carries a risk of injury if the collimator comes into contact with the technologist or the patient. Recall start date: November 26, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.