Official recall record
Azurion 7 M20 (ref 74842)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-01-04
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has become aware of two potential safety issues with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software version 05.02.12 ("table"). Due to communication issues between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. There can be a connection issue (software issue with table to be corrected in field action FCO72200557). There is also a date and time issue where when the flexarm system is (re)started, the flexarm and table may not connect due to a software issue. This occurs when the date/time format includes more than three (3) leading zeros (to be corrected in field action FCO72200528). Recall start date: December 13, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.