Official recall record
Azurion 7 M20 (ref 74843)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-01-04
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has become aware of a potential safety issue with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software versions 05.02.12 and 05.02.15 ("table"). Due to a communication issue between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. A representative of Getinge (manufacturer of the Maquet Magnus Table) will arrange a site visit to return the software of your table to version 05.02.12. After this change, your system will be subject to the medical device recall letter (HS23039R/2023-IGT-BST-020). Recall start date: December 13, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.