Official recall record
Azurion 7 System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-01-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has become aware of a potential safety issue with the Philips Azurion Systems R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2, where the system may exhibit a loss of imaging functionality and data. Philips has identified a potential issue which can cause the system to continuously restart (restart loop). T his is likely to occur when the patient database gets too big over time (>500 studies). If this issue occurs, a potential for data loss may be expected. Recall start date: December 15, 2023