Official recall record
Pulse*Spray® Infusion Systems 3fx135x120
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-01-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Angiodynamics, Inc., is conducting a medical device recall to the end user level based on information received from ICU medical (Smith Medical), the manufacturer of the non-sterile high-pressure rotator which are components of the sterile Angiodynamics Pulse*Spray® Infusion Catheter Systems. ICU medical has determined that specific non-sterile high-pressure rotator lots, listed in their supplier notification dated November 16, 2023, contain a manufacture defect. As stated in the information provided by ICU medical, "The inner diameter of the device's [rotating adaptor's] o-ring may be oversized affecting seal integrity." Angiodynamics has confirmed that the affected Pulse*Spray® Infusion Systems were distributed to end users located in the United States, Europe, and Canada. Angiodynamics began distributing Pulse*Spray® Infusion Systems affected by this recall globally on August 29, 2023. Angiodynamics has not received any complaints associated with the affected Pulse*Spray® Infusion Systems or reports of patient injury as a result of this issue. Recall start date: December 18, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.