ProductGuru Recalls

Official recall record

Pulse*Spray® Infusion Systems 3fx135x120

Health Canada — Medical Devices · 2024-01-05 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-01-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Angiodynamics, Inc., is conducting a medical device recall to the end user level based on information received from ICU medical (Smith Medical), the manufacturer of the non-sterile high-pressure rotator which are components of the sterile Angiodynamics Pulse*Spray® Infusion Catheter Systems. ICU medical has determined that specific non-sterile high-pressure rotator lots, listed in their supplier notification dated November 16, 2023, contain a manufacture defect. As stated in the information provided by ICU medical, "The inner diameter of the device's [rotating adaptor's] o-ring may be oversized affecting seal integrity." Angiodynamics has confirmed that the affected Pulse*Spray® Infusion Systems were distributed to end users located in the United States, Europe, and Canada. Angiodynamics began distributing Pulse*Spray® Infusion Systems affected by this recall globally on August 29, 2023. Angiodynamics has not received any complaints associated with the affected Pulse*Spray® Infusion Systems or reports of patient injury as a result of this issue. Recall start date: December 18, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.