Alinity I HBsAg Reagent Kit

Health Canada — Medical Devices · 2024-01-08 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-01-08
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Abbott has identified a few cartridges within Alinity I HBsAg reagent kit, list number 08P0852, lot 51503FN00 that may exhibit variability in relative light unit (RLU) response and concentration values which may result in controls out of range, and/or incorrect patient results. Positive control out of range low results and/or negative control out of range high results could occur when using an impacted cartridge. Per the instructions for use (IFU), controls are required to be run once within each 24 hours of use. A potential falsely decreased or falsely increased patient result may occur for any of the following scenarios: 1. An impacted cartridge was used without running controls. 2. An impacted cartridge was used following calibration with a non-impacted cartridge. 3. An impacted cartridge was successfully calibrated, and controls were within range but calibrator and/or control RLUs are low. Recall start date: December 21, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.