Official recall record
Rotaflow Centrifugal Pump With Bioline
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-01-09
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
During retrospective internal testing, MCP/Getinge discovered narrow channels in the seal of the sterile bags used to package the RF-32. MCP/Getinge receives this sterile bag from a supplier that performs tests to verify sterility. Incomplete sealing can potentially lead to a breach of the sterile barrier. Recall start date: December 22, 2023