What the authority reported
The quadrature body coil (QBC) seal adhesive may fail creating sharp edges that may come in contact with patients. The QBC seal may become loose as the patient table travels in a horizontal motion in and out of the system bore. The QBC seal (figure 1) is a rubber seal that is glued between the cone cover and QBC cover and functions to prevent sharp edges of the QBC cover from contacting patients during an examination. Recall start date: January 9, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.