Official recall record
Ionostar Plus and Ionolux
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-02-02
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionostar Plus. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the Instructions For Use (IFU) is not possible. The reactive component in the incorrect plungers contains small amounts of BHT and Methacrylate. These ingredients are not listed in the IFU. Recall Start Date: January 23, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.