ProductGuru Recalls

Official recall record

Olympus Endoscope Reprocessor

Health Canada — Medical Devices · 2024-02-08 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-02-08
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

This recall pertains to the MAJ-2318 water filter, which is compatible with the following Olympus endoscope reprocessors: OER-Elite. This water filter has been designed to prevent rinse water contamination in the OER endoscope reprocessors. Olympus confirmed there is a possibility that the water filter MAJ-2318 may have been assembled improperly. Specifically, the adapter may be installed on the incorrect end of the filter body. If customers use a defective filter, the tap water supplied to the OER endoscope reprocessors may not be properly filtered. As a result, unfiltered tap water would be used for reprocessing. Recall start date: January 22, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.