Olympus Endoscope Reprocessor
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-02-08
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
This recall pertains to the MAJ-2318 water filter, which is compatible with the following Olympus endoscope reprocessors: OER-Elite. This water filter has been designed to prevent rinse water contamination in the OER endoscope reprocessors. Olympus confirmed there is a possibility that the water filter MAJ-2318 may have been assembled improperly. Specifically, the adapter may be installed on the incorrect end of the filter body. If customers use a defective filter, the tap water supplied to the OER endoscope reprocessors may not be properly filtered. As a result, unfiltered tap water would be used for reprocessing. Recall start date: January 22, 2024