ProductGuru Recalls

Official recall record

VITROS® Systems (ref 75059)

Health Canada — Medical Devices · 2024-02-09 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-02-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Issue #1 - QC baseline statistics are not updating as expected. This only affects VITROS® Systems running VITROS® system software version 3.8.1. Issue #2 - Configurable westgard rules are not being applied as expected. This affects VITROS® Systems running VITROS® System software versions 3.8.0 and 3.8.1. Recall start date: January 31, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters