Official recall record
Panther System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-02-13
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
According to the assay package inserts for the following Aptima assays: HIV-1 Quant DX, HBV Quant, HCV Quant DX, CMV Quant, BV, and CV/TV, controls from different lot numbers can be used. However, Hologic has identified a software anomaly with Panther/Fusion SW version 7.2.9, where assay processing will not start when the control lots loaded on the Panther/Panther Fusion system do not match the ones listed on the control barcode sheet. Recall start date: November 17, 2023
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.