ProductGuru Recalls

Official recall record

Ambu Aview 2 Advance

Health Canada — Medical Devices · 2024-02-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-02-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Ambu have received input from market where Ambu® Aview 2 advance was dropped and upon impact hitting the floor the lithium-ion battery in the device short-circuited and combusted whereby smoke and flames was seen. No harm to patient, user or third party has been reported in any of the cases. Recall start date: August 13, 2023

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.