ProductGuru Recalls

Official recall record

Care Plus Incubator

Health Canada — Medical Devices · 2024-02-15 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-02-15
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

GE Healthcare has become aware that the bedside panel of the incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling. Recall start date: February 5, 2024