SOLTIVE™ Premium SuperPulsed TF Laser System

Health Canada — Medical Devices · 2024-02-29 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-02-29
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

As part of Olympus' post market surveillance of SOLTIVE™ laser use, including a review of customer complaints and adverse events, we have become aware that some clinicians are defaulting to the preset laser values for all clinical procedures when that is not consistent with the instructions for use. Olympus is reminding users that preset treatment parameters may not be appropriate for all patients and are guidelines only. Olympus is sending this reminder after reviewing twenty-eight (28) complaints reported between 2021 to 2023 related to ureteral stone procedures, that described serious injuries. Recall start date: February 5, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters