BD Bodyguard Microset

Health Canada — Medical Devices · 2024-03-06 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2024-03-06
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

BD received customers complaints, that the new clamp on the y-port of the BD Bodyguard™ Microset (clamp supplied since 2021) can be opened more easily compared to the clamp which was used before 2021. Patients may accidentally lay on the infusion set and unintentionally open the clamp causing the infusion set to be exposed to atmospheric air/ environment, if a vented clamp is being used. The longer this failure mode remains undetected the higher the potential volume of air entry. Recall start date: February 26, 2024