Official recall record
Origio Sperm Wash
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-03-07
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount). Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium. Recall start date: January 22, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.