Digital Radiography Mobile X-Ray System Dx-D 100
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-03-12
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
A component supplier of the Agfa Dx-D 100 informed Agfa about a potential safety issue. The issue as described by the supplier is related to the steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm if the cable is cut off does not work properly. The interference occurs between the steel cable and a metal plate of the arm's carriage due to the shape of the cable's end. Recall Start Date: February 23, 2024