Patient Information Center iX (ref 75241)

Health Canada — Medical Devices · 2024-03-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-03-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

It was found that the care assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when ios device screens were locked with a passcode. The issue occurred on devices running IOS 15, 16 and 17 as well as multiple versions of iPhones. The care assist mobile application is not intended to replace the patient's primary alarming system. If users rely on the care assist application as their primary event notification alarming system, the failure of custom ringtones could result in a delay in treatment for patients experiencing a change in clinical condition. Recall Start Date: March 6, 2024