ProductGuru Recalls

Official recall record

Medfusion Syringe Infusion Pumps

Health Canada — Medical Devices · 2024-03-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-03-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Smiths Medical is issuing this letter to notify customers not to use Cardinal Health branded Monoject syringes with Medfusion 3500 and 4000 Syringe Pumps. On September 20, 2023, Cardinal Health issued a medical device product correction for the Cardinal Health branded Monoject Luer-Lock Tip Syringes due to demonstrated recognition and compatibility issues with syringe infusion pumps due to changes made to the dimensions of the syringes. As a result, Cardinal Health recommends that these syringes not be used on syringe infusion pumps. Recall Start Date: February 29, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.