Diamed ID-Systems

Health Canada — Medical Devices · 2024-03-22 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-03-22
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An increased level of complaints and adverse events related to weak non-specific reactions when using ID-system screening reagents in Indirect Antiglobulin Testing (IAT) for qc, donor, and patient samples. Customers using both, the manual and instrument methods reported weak positive reaction results (identified as "?", "wr", "+/-", and sometimes "+" on instruments) instead of an expected clearly negative "-" reaction in IAT. These doubtful reactions or in some cases weak positive reactions may occur randomly among batches. The investigations so far point out that a possible enhancer of the phenomenon is the combination of certain lots of reagents red blood cells with certain lots of associated ID-cards. Recall start date: March 11, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.