Siemens Biograph mMR System

Health Canada — Medical Devices · 2024-03-25 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2024-03-25
Last updated (per notice)
2024-03-22
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Siemens Healthcare GmbH became aware of this issue as part of a customer complaint investigation. It has been identified that this issue can occur when certain parameters (Echo spacings 0.50 MS; 0.51 MS, and 0.52 MS) are used in EPI (Echo-Planar Imaging), sequences will excite high internal stress inside the gradient coil assembly. This may lead to cracks and damage to the gradient coil. Recall start date: March 15, 2024