Official recall record
Siemens Biograph mMR System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-03-25
- Last updated (per notice)
- 2024-03-22
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Siemens Healthcare GmbH became aware of this issue as part of a customer complaint investigation. It has been identified that this issue can occur when certain parameters (Echo spacings 0.50 MS; 0.51 MS, and 0.52 MS) are used in EPI (Echo-Planar Imaging), sequences will excite high internal stress inside the gradient coil assembly. This may lead to cracks and damage to the gradient coil. Recall start date: March 15, 2024