What the authority reported

CHS initiated this recall after receiving a notification from Terumo on their cardiovascular procedure kits that include high crack check valve. Terumo cardiovascular systems (Terumo CVS) has received six complaints whereby the pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmhg versus 400mmhg. Recall start date: February 26, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →