Official recall record
Terumo Cardiovascular Procedure Kits
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-03-28
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
CHS initiated this recall after receiving a notification from Terumo on their cardiovascular procedure kits that include high crack check valve. Terumo cardiovascular systems (Terumo CVS) has received six complaints whereby the pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmhg versus 400mmhg. Recall start date: February 26, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.