Official recall record
Prelude Roadster Guide Sheath
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2024-04-08
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Merit Medical Systems, Inc. is voluntarily conducting a recall of the Prelude Roadster Guide Sheath due to an incorrect dilator included with the product. This is limited to sheaths labeled with .018" dilators, which may contain .038" dilators. This affects two lots. Merit has not received any reports of patient harm or injury relating to this issue; however, Merit has received one complaint relating to this issue. Use of the incorrect dilator may result in procedure delay or soft tissue injury. Recall Start Date: March 14, 2024
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.